- Pin BB 0123456789
- WhatsApp 1234567890
- Line 2345678901
Detail Produk
Model:BT-XC04
Usage:
Orthopedics: restore bone translocation, reset, fixing
Surgery: taking foreign bodies out of the body, cardiac catheterization, implantable pacemakers, interventional treatment, some of angiography and local photography etc.
Features:
- New integrated design of TV system and mainframe, it does not need to connect electric cable, so the operation is more convenient.
- High-quality combined HF-HV X-ray generator with our high voltage cable, it reduces loss and the quality of x-ray and image are good.
- Electric rotating lead piece iris beam limiting device, it can satisfy multi-angle and multi-direction anatomical display.
- The machine adopts imported image intensifier, matching with high-definition and high-resolution CCD and progressive scan display system, which matches with image system better.
- With the perspective KV, MA automatic tracking feature, the image brightness and definition are automatically set at optimal function.
- Graphical colorfully LCD touchable screen console interface, the operation is convenient and intelligent.
- Unique hand-held remote controller, the doctor can operate it a far distance, so it’s more convenient.
Configurations:
- New integrated C-arm mainframe one set
- Combined high-frequency high-voltage X-ray generator and high-frequency power supply
(3.5kW, 40 kHz, 110kV) one set
- imported 9 inch image intensifier one set
- Imported ultra-low illumination CCD radiography one
- High-definition progressive output CCU control system one set
- High-definition progressive output CCU control system one set
- Imported dense grain grids one set
- Electrical adjustable beam limiting device one set
8.14-inch high-definition medical dedicated monitor one
- Unique hand-held controller one set
Performance and Parameter:
Category | Items | Content |
Electrical performance | High frequency inverter power supply | Power:3.5KW Main inverter frequency: 40 kHz |
Perspective max rated capacity for fluoroscope | Tube current 8 mA Tube voltage 110 kv | |
Automatic fluoroscopy | tube voltage 40 kv~110kv,adjust automatically
tube current : 0.3mA~~4mA adjust automatically |
|
Manual fluoroscopy | tube voltage 40 kv~110kv, continuous
tube current : 0.3mA~~4mA continuous |
|
Pulse fluoroscopy | tube voltage 40 kv~110kv, continuous
tube current : 4.1mA~~8mA continuous |
|
Photography tube voltage, tube current mAs | 40KV~110KV, 20~63mA, 1~125 mAs | |
Plate holder size | 200mmx250mm(8”x10”) or 250mmx300mm(10”x12”) | |
X-ray tube | X-ray tube special for high frequency | large focus:1.5, mall focus :0.3
Anode thermal capacity: 35kJ (47 kHu) Tube thermal capacity: 650kJ (867 kHu) |
Imaging
system |
Image intensifier | Imported 9 inch image intensifier |
CCD photography | Imported ultra-low illumination CCD radiography | |
Monitor | Center definition is 1000 lines and frequency for picture refresh is 75Hz. | |
CCU (central control) | HD progressive output, continuous adjustable recursion, many images storage, upright image, horizontal image, positive & negative image, image patching, LIH (last image freeze), OSD (monitor display). | |
Structure and performance | Direction wheel and main wheel | Direction wheel can rotate in any direction, and main wheel can rotate in ±90°. |
C-arm | The up and down electrical stroke of pillar is 400mm.
Monitor rotating angles ≥300° Forward and backward movement:200mm; Revolution around horizontal axis : ±180°; Revolution around vertical axis: ±15°, distance from focus to screen: 960 mm; C-arm open distance: 760mm C-arm arc depth: 640mm; Slipping on orbit : 120°(+90°~-30°) |