Detail Produk
-Microprocessor and high speed servo controlled,volume and pressure mode system.
–SIMV mode with super sensitivity flow-trigger.
–Compliance and leakage compensation system.
–P-V loop & F-V loop, dynamic compliance.
–10.4’TFT screen display.
–Self testing,trouble shooting and warning system
–High quality vaporizer
–Low resistance,high precise and compact absorber system with heating function.
–Big screen Monitor and AMIS(Anesthesia Management Information System) .
–Mode: Semi-open, semi-closed or closed system–Gas supply: O2/Air/N2O, supply pressure: 0.3 – 0.5Mpa–Flowmeter: O2 / Air / N2O Five-tube Flowmeter. 0-10L(With inter-lock protection devices ensuring the oxygen concentration above 25%).–O2 flush: 35L/min – 75L/min–O2 deficiency: Audible alarm is long for at least 7 seconds when the O2 supply downp
with top-light system
–Canister prompt replacing and absorber heating system
Power supply | 100V ÷ 240V; 50Hz / 60Hz |
---|---|
Battery backup | ≥ 40 minutes |
Gas supply | O2, N2O, Air; pressure 280 ÷ 600kPa |
Ventilator | Electronically controlled, pneumatically driven |
Ventilation operating modes | Standby, VCV, PLV, SIMV, Manual/Spontaneous breath |
Pressure range | 5 ÷ 70cm H2O |
Tidal volume | 50 ÷ 1500ml |
Breathing frequency | 4 ÷ 100bpm |
Inspiration/Expiration ratio | 1:0,5 to 1:8 |
Inspiratory plateau | 5 ÷ 50% |
Positive end-expiratory pressure | 5 ÷ 20cm H2O |
High pressure | 20÷ 80cm H2O |
Low pressure | 0 ÷ 20cm H2O |
Trigger sensitivity | 1 ÷ 30L/minute |
Parameters monitored | VT , MV, Paw (Peak, Pplat), FiO2, breath rate, compliance |
Waveforms monitored | P-T, F-T, V-T waveforms, P-V, F-V loops |
Display | 10,4” TFT color screen |
Setting | touch key & navigator wheel knob |
Patient mode | Adult & pediatric |
Mode | close, semi-close or semi-open |
Fresh gas flow indicators | O2 = 0,05 ÷ 1; 1 ÷ 10L/minute N2O = 0,05 ÷ 1; 1 ÷ 10L /minute AIR = 0,1 ÷ 10L/minute |
Integrated safety functions | Oxygen Ratio Controller (ORC) guarantees O2 concentration of >25% with anti-hypoxia system N2O cut-off if O2 fresh gas valve is closed or if O2 flow is less than 50%. |
O2 flush | 35 ÷ 75L/minute |
Safety valve | ≤ 6kPa |
Two vaporizer mount | Selectatec system; Interlock |
Alarms | high/low limit of breath rate, Paw high/low limit, MV high/low limit, low O2 concentration, airway malfunction, power failure |