SPESIFIKASI :

  • Product Name: TERUFUSION Syringe Pump Terumo SS 700
  • Compatible Syringes 10, 20, 30 and 50 mL (TERUMO and specified brands)
  • Range of Flow Rate Setting:
    • 0.1-300 mL/h (for 10, 20 and 30 mL syringes)
    • 0.1-999 mL/h (for 50 mL syringes)
    • 0.1-1200 mL/h (special function for 50 mL syringes)
  • Flow Rate Display Change:
    • 0.1 mL/h steps (for settings from 0.1 to 99.9
    • mL/h) 1 mL/h steps (for settings from 100 to 999 mL/h)
  • Flow Rate Accuracy:
    • Mechanical Accuracy: ±1%
    • Accuracy including syringe: ± 3%
  • Occlusion Detection Pressure: Selectable from three levels. (Values are with use of TERUMO syringe)
    • [III]: 800 ± 200 mmHg max. (106.7 kPa ± 26.7 kPa)
    • [II]: 500 ± 100 mmHg max. (66.7 kPa ± 13.3 kPa)
    • [I]: 300 ± 100 mmHg max. (40 kPa ± 13.3 kPa)
    • Optional: 5 occlusion levels (100, 200, 300, 500 and 800 mmHg)
  • Priming (purge rate): 300 mL/h ~ 1200 mL/h (Depending on syringe size)
  • Alarms: Occlusion, Nearly empty, Low Battery, Plunger/Clutch disengaged, Syringe disengaged, AC/DC Cable disconnected. High flow warning when turning the dial over 10 mL/h (optional)
  • Special Functions: Functions are selectable via internal switches:
    • Delivery Limit setting:
      • 0.1 – 999.9 mL (in 0.1 mL steps)
      • When the limit is reached, an alarm will occur and the infusion will continue at KVO rate of 0.1 mL/h.
    • Nearly Empty alarm setting:
      • The time until the syringe is empty can be programmed. (from 3 to 60 min.)
    • Max. Flow Rate:
      • 1 – 1200 mL/h (in 1 mL steps)
    • Syringe Brand Confirmation
    • Memory function (last settings)
    • Changeable flow rate during infusion
  • Power Supply:
    • 100 to 240 VAC, 50/60 Hz
    • 12 to 15 VDC
    • Operable duration using internal battery NiMH:
    • Approx. 5 hours (using a new battery at flow rate of 5 mL/h and ambient temperature of 25°C, after charging at least 15 hours with the power off)
  • External Dimensions: 322 (W) x 114 (H) x 115 (D) mm
  • Weight: TE-331 ± 1.8 kg

Produk Pernah Dilihat

Detail Produk

Kategori : LAIN-LAIN

SPESIFIKASI :

  • Product Name: TERUFUSION Syringe Pump Terumo SS 700
  • Compatible Syringes 10, 20, 30 and 50 mL (TERUMO and specified brands)
  • Range of Flow Rate Setting:
    • 0.1-300 mL/h (for 10, 20 and 30 mL syringes)
    • 0.1-999 mL/h (for 50 mL syringes)
    • 0.1-1200 mL/h (special function for 50 mL syringes)
  • Flow Rate Display Change:
    • 0.1 mL/h steps (for settings from 0.1 to 99.9
    • mL/h) 1 mL/h steps (for settings from 100 to 999 mL/h)
  • Flow Rate Accuracy:
    • Mechanical Accuracy: ±1%
    • Accuracy including syringe: ± 3%
  • Occlusion Detection Pressure: Selectable from three levels. (Values are with use of TERUMO syringe)
    • [III]: 800 ± 200 mmHg max. (106.7 kPa ± 26.7 kPa)
    • [II]: 500 ± 100 mmHg max. (66.7 kPa ± 13.3 kPa)
    • [I]: 300 ± 100 mmHg max. (40 kPa ± 13.3 kPa)
    • Optional: 5 occlusion levels (100, 200, 300, 500 and 800 mmHg)
  • Priming (purge rate): 300 mL/h ~ 1200 mL/h (Depending on syringe size)
  • Alarms: Occlusion, Nearly empty, Low Battery, Plunger/Clutch disengaged, Syringe disengaged, AC/DC Cable disconnected. High flow warning when turning the dial over 10 mL/h (optional)
  • Special Functions: Functions are selectable via internal switches:
    • Delivery Limit setting:
      • 0.1 – 999.9 mL (in 0.1 mL steps)
      • When the limit is reached, an alarm will occur and the infusion will continue at KVO rate of 0.1 mL/h.
    • Nearly Empty alarm setting:
      • The time until the syringe is empty can be programmed. (from 3 to 60 min.)
    • Max. Flow Rate:
      • 1 – 1200 mL/h (in 1 mL steps)
    • Syringe Brand Confirmation
    • Memory function (last settings)
    • Changeable flow rate during infusion
  • Power Supply:
    • 100 to 240 VAC, 50/60 Hz
    • 12 to 15 VDC
    • Operable duration using internal battery NiMH:
    • Approx. 5 hours (using a new battery at flow rate of 5 mL/h and ambient temperature of 25°C, after charging at least 15 hours with the power off)
  • External Dimensions: 322 (W) x 114 (H) x 115 (D) mm
  • Weight: TE-331 ± 1.8 kg